Model 2231A Triple-channel DC Power Supply Reference Manual Section 5: SCPI command reference
077100401/September 2014 5-27
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]
This command configures voltage stepping.
Item Description
Group Source
Syntax [SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement] <voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate]:STEP[:INCRement]?
Arguments The voltage step for UP/DOWN adjustment
Unit kV|V|mV|uV
Returns None
Examples Volt:step 0.01
Volt:step?
[SOURce:]VOLTage[:LEVel]:TRIGgered[:IMMediate][:INCRement]
This command is to set the level for the trigger function. When you set the trigger voltage and current and send
the trigger command to the instrument, the voltage and current can switch to the trigger voltage and current that
you set before.
Item Description
Group Source
Syntax [SOURce:]VOLTage[:LEVel]:TRIGgered[:IMMediate][:INCRement]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:IMMediate][:INCRement]?
Arguments 0 to max rated voltage
Unit kV|V|mV|uV
Returns Trigger voltage
Examples Volt:trig 12
[SOURce:]VOLTage:LIMIT[:LEVel]
This command is to set the voltage limit for the present channel.
Item Description
Group Source
Syntax [SOURce:]VOLTage:LIMIT[:LEVel]
[SOURce:]VOLTage:LIMIT[:LEVel]?
Arguments 0 to max rated voltage||Min|Max|def
Returns The voltage limit