Section 11: TSP command reference Series 3700A System Switch/Multimeter Reference Manual
11-236 3700AS-901-01 Rev. D/June 2018
errorqueue.next()
This function reads the oldest entry from the error queue and removes it from the queue.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
errorCode, message, severity, errorNode = errorqueue.next()
errorCode
The error code number for the entry
message
The message that describes the error code
severity
The severity level (0, 10, 20, 30, or 40); see Details for more information
errorNode
The node number where the error originated
Details
Entries are stored in a first-in, first-out (FIFO) queue. This function reads the oldest entry and
removes it from the queue.
Error codes and messages are listed in the Error summary list (on page 12-9).
If there are no entries in the queue, code 0, "Queue is Empty" is returned.
Returned severity levels are described in the following table.
Severity level descriptions
Number
Level
Description
0
Informational
Indicates that there are no entries in the queue.
10
Informational
Indicates a status message or minor error.
20
Recoverable
Indicates possible invalid user input; operation continues but action
should be taken to correct the error.
30
Serious
Indicates a serious error that may require technical assistance, such as
corrupted data.
40
Fatal
Instrument is not operational.
In an expanded system, each TSP-Link enabled instrument is assigned a node number. The variable
errorNode stores the node number where the error originated.
Example
errorcode, message = errorqueue.next()
print(errorcode, message)
Reads the oldest entry in the error queue. The
output below indicates that the queue is empty.
Output:
0.00000e+00 Queue Is Empty
Also see
Error Available Bit (Error or Event queue) (on page C-8)
errorqueue.clear() (on page 11-235)
errorqueue.count (on page 11-235)
Error summary list (on page 12-9)

Table of Contents

Other manuals for Keithley 3730

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3730 and is the answer not in the manual?

Keithley 3730 Specifications

General IconGeneral
BrandKeithley
Model3730
CategoryMultimeter
LanguageEnglish

Related product manuals