75 PBS/PBJ-BA-e-1718
Error message:
Displayed error
code
Explication
Remedy
Disturbances in weight loading
mechanical elements.
Check transport screws.
Weight on the scale pan is unstable
during calibration.
Large zero point drift during
calibration.
Empty the scale pan.
Large drift during the PCAL function.
Use the correct weight.
Large drift during calibration of the
measuring range.
Use the correct weight.
Calibration weight is incorrect.
Use the correct weight.
Disturbances in the balance (the
balance is stopped when this display
appears).
Received command code is
incorrect.
Check limiters etc.
Total number of the displayed unit is
longer than 7 positions.
Reduce load.
Disturbance in the balance.
The power voltage is incorrect.
Check the power voltage .