PBS/PBJ-BA-e-1718 76
Fault finding Procedures:
• Mains adapter is not connected.
• The room mains circuit-breaker is
switched off.
connect the mains adapter
correctly.
The transport screws are not locked.
Turn the screws anticlockwise
until they are locked.
Pads of the scale pan not installed.
Install pads of the scale pan.
To large weight on the scale pan.
Use the weight within its capacity.
react after putting a
weight on the scale
The weighing pan is displaced.
Place the weighing pan correctly.
fluctuates.
Influence of vibrations or air flow.
Place the weight at the suitable
location.
Try to change environment
Protective shield is touching the
scale pan.
Install the shield on the main unit
of the balance.
Incorrect weighing
result.
Calibration of the measuring range is
not effective.
Calibrate the balance correctly
Taring is not effective.
Tare the balance before
weighing.
display the required
The unit has not been set.
Selection of the menu
element is rejected.