7.3 Assembling, Installation and Levelling
The correct implantation site is important for the accuracy of the weighing
results of high-resolving precision balances (see chap. 7.1).
7.3.1 Implantation of the balance
1. Place the weighing plate support of the balance (in PES 31000-1M the
weighing plate support is pre-installed)
2. Fix the weighing plate support with the screw
3. Put the weighing plate on the weighing plate support
4. Level balance with foot screws until the air bubble of the water balance is in
the prescribed circle
Check levelling regularly
5. Connect the mains adapter (Installation of the mains adapter: see chap.
7.4.1)