Error message
Description
Possible causes / repair
Maximum weighing range exceeded
Split sample and weigh
individually
Use lighter tare weight
Negative load is below the minimum
weighing range
Weighing plate or
weighing plate carrier
incorrectly adjusted
Check whether the
balance is touching other
objects
Weight value of the sample when
setting the reference weight in piece
counting mode is too low
Use samples / reference
weights with higher weight
value (lowest piece
weight, minimum load)
The [S] key was pressed, although
<*> has not been displayed
Observe the totalizing
procedure according to the
Operating instructions
System error
Inform the retailer.
The weight value of the adjustment
weight is less than 50 % of the
weighing capacity.
The external adjustment weight is
less than 95 % of the weighing range
when calibrating the internal
adjustment weight
Use a adjustment weight
with a weight value as
close as possible to the
weighing capacity.
Error > 1.0 % at adjustment test with
external weight
Weighing plate is loaded during the
internal adjustment
Unload the weighing plate
and repeat the internal
adjustment
Error > 1.0 % at the internal
adjustment
Perform internal
adjustment again
The input value for the measurement
incertainty of the external adjustment
weight at <2. o.M.P.> exceeds the
maximum setting range of +/- 100 mg
Use adjustment weights
with poor deviation
Faulty end of the internal adjustment
Perform internal
adjustment again

Table of Contents

Other manuals for KERN PES 620-3M

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN PES 620-3M and is the answer not in the manual?

KERN PES 620-3M Specifications

General IconGeneral
BrandKERN
ModelPES 620-3M
CategoryScales
LanguageEnglish

Related product manuals