2− 4
Section 2 Installation and Power Supply
㸰㸫㸱
Place of Installation
㸰㸫㸲
Movement of Centrifuge
CAUTION
●
Center of gravity position is around a center of centrifuge.
●
When moving the centrifuge, lift up the bottom from the sidepiece of the centrifuge
body by two or more persons using equal forces to move the centrifuge to the
desired place.
●
Do not drop the centrifuge, otherwise damage or injury may occur.
CAUTION
Do not install the centrifuge on an inclined, slippery, or unstable surface.
Violent vibration may occur.
WARNING
A clearance of 30 cm minimum must be provided around the centrifuge.
If the centrifuge is driven into uncontrolled rotations due to a failure, secondary
damage can result from energy absorbed by the rotation.
WARNING
●
Never move the centrifuge while the rotor is rotating or while the rotor is
attached to the centrifuge. Otherwise, the drive shaft may become bent or the
URWRUPD\FRPHRႇUHVXOWLQJLQDQDFFLGHQWRUGDPDJHWRWKHFHQWULIXJH
●
Ensure that rotor are removed from the centrifuge and that the power supply cable
is disconnected from the wall socket.
●
Moving the position of the centrifuge while the power is turned on may cause
HOHFWUL¿FDWLRQDFFLGHQWRUIXQFWLRQDOIDLOXUHRIWKHFHQWULIXJH