3− 12
Section 3 Operation
100rpm
min
sec
TIME
min
sec
TIME
㹑㹒㹍㹎
㹌
㹰㹮㹫
SPEED
100rpm
SPEED
㹙
1
㹛
Setting the "Natural" Deceleration
Thecharacteristicshavebecomeasshowninthegraph.
Youcansetthenaturaldecelerationstartingspeed
N
bythefollowingmethod.
Factory default
,
it is set as 0 rpm
.
The setting range of “ N ” is from 0 to 12,000 rpm.
(1)Pressingtheknobwhileholdingdownthe"STOP"key.
Theusednumberofcyclesofthecentrifugeisdisplayed.
(2)Pressingtheknobseveraltimesandletindicate
"Fr"
onthe"TIME"display.
The"SPEED"displayindicatesthecurrentsettingofthenaturaldecelerationspeed.
▶
Refer to page 3-11.
(3)Turntheknobtosetthenaturaldecelerationspeed.
(4)Presstheknobwhileholdingdownthe"STOP"key.
ThedisplayreturnstothepreviousSpeedandTimeindicationafterthesetting(Natural
decelerationspeed)issaved.
㹙
Ex.
㹛ࠉ
Setting is the 0 rpm.
㹙
Ex.
㹛
Setting the 300 rpm.
speed
RAPID
TIME
Natural
deceleration