2− 5
Section 2 Installation and Power Supply
㸰㸫㸳
Power Requirements
WARNING
Prepare a power supply that meets the following conditions.
1. The power supply voltage must be the same as that indicated on the nameplate of
the centrifuge and the voltage variation must be within the range given in Table
2-1.
2. The current capacity must be more than given in Table 2-1.
3. In connecting the cord, install a knife switch or circuit breaker of given value in
Table 2-1.
When the centrifuge is connected to an outlet by means of a plug, use a 3-pin plug
with a ground terminal.
4. Use single-phase power.
5. The outlet must have a ground terminal and its ground resistance must be less
than 100 ohm.
RatedVoltage 110V 115V 220V 230V
AcceptableVoltageRange 99Vto121V 103Vto126V 198Vto242V 207Vto253V
Currentrequirement 10A 10A 6A 6A
WARNING
●
The power cable alone should be connected to the plug socket.
●
'RQRWXVHDEUDQFKHGSOXJVRFNHWZKLFKPD\FDXVHRYHUKHDWLQJRU¿UH
Table 2-1 Rated Voltage
,
Current and Acceptable Voltage Range and current requirement