4− 1
Section 4 Service
Section 4
Service
㸲㸫㸯㸬
Daily Inspection
WARNING
If any abnormality is found during routine daily inspections, discontinue centrifuge
operation, use the "POWER" switch to turn the machine OFF, disconnect the power
cable, attach "DO NOT USE" labels to the rotor and the outside of the centrifuge,
and contact your nearest dealer to request a centrifuge inspection before resuming
operation of the equipment.
If the centrifuge is used after discovering any abnormality, an accident could occur,
possibly leading to serious damage or accidents involving physical injury.
Checkpoint Action taken
Check whether the rotor cover
and rotor locking bolt are properly
tight.
Retighten,ifeitheroftheseisloose.Fordetails,refer
to
Page 2-3.
Check carefully for any cracks,
evidence of corrosion, rust or
deformation on the rotor and
metal tubes.
Donotusetherotorandmetaltubesifanycracks,corrosion,
rustordeformationsarefound.Contactyournearestdealerfor
aninspection.
Check that no foreign material
or water is present in the
chamber.
Removeanyforeignmaterialorwaterbeforeoperatingthe
centrifuge.
Check that the hook is locked
completely.
Unlessthelatchislockedcompletely,thelidmayopenduring
centrifugeoperation.
Stopcentrifugeoperationimmediatelyandcontactyour
nearestdealerforaninspection.
&KHFNWKDWWKHVFUHZV¿[LQJWKH
lid hinges are not loose.
Ifanyofthesesecuringscrewsarelooseorhavebeen
removed,stopcentrifugeoperationimmediatelyandcontact
yournearestdealerforaninspection.
Check that the grounding wire is
correctly connected.
▶
Refer to Page 2-6.
Check that all knobs, displays,
lamps and switches operate
correctly.
Ifanydonotoperatecorrectly,contactyournearestdealerfor
aninspection.
▶
Refer to Page 1-1 and 1-2.
Check that the centrifuge is
properly leveled.
Ifitisnotproperlyleveled,carryoutappropriateadjustments.
▶
Refer to Page 2-4.