D9U001PPB-010132/106
Permitted back section load (125 mm from the edge) 150 kg (330 lb)
Permitted foot section load (125 mm from the edge) 150 kg (330 lb)
Permitted leg rest load 70 kg (154 lb)
Permitted load of side rail 75 kg (165 lb) vertical press
Permitted load of supporting bar 70 kg (154 lb)
Weight of mattress
(back and seat section 100/130mm, 2.56-5.11 inch)
8.6 - 9.6 kg (19-21 lb)
Weight of mattress (foot section) 1.2 kg (2.64 lb)
Weight of mattress (pair of leg rest) 2.6 kg (5.73 lb)
Noise level Less than 65 dB (A)
7.3 Electrical parameters
Batteries – back-up 24V/2,3 Ah
Voltage – input 220 - 240 VAC
Voltage – input 110 - 120 VAC
Voltage – input 100 VAC
Voltage – input 127 VAC
Frequency 50/60 Hz
Nominal value of the fuse in the control unit
(220-240V)
T 3,15 AL, 250V
Nominal value of the fuse in the control unit (100 V) T 6,3 AL, 250V
Nominal value of the fuse in the control unit(110-120V) T 6,3 AL, 250V
Nominal value of the fuse in the control unit (127V) T 6,3 AL, 250V
Maximum input 570 VA
Motor voltage 24 V
Ingress protection of the birthing bed IPX4
Device class I.
Classification of applied parts B
All possibilities of voltage – input 100 VAC 50/60 Hz
110-120 VAC 50/60 Hz
127 VAC 50/60 Hz
220-240 VAC 50/60 Hz
Functional description of motors:
The birthing bed is fitted with electrical linear motors for moving the back section, seat section and
for bed height adjustment. These motors require no maintenance.