D9U001PPB-0101 45/106
9.3 Manual setting of the CPR
WARNING!
The CPR controller is marked with red information labels on both sides of the
back section. The labels are marked with CPR text
WARNING!
During movement, the back support must be loaded (e.g. with the
patient's weight)
NOTE Before first use of the bed, we recommend performing the movement of the back rest of the
bed multiple times using the CPR function. The CPR function may be slightly impaired if the
bed has not been used for some time
IMPORTANT To start the CPR function, the patient must be on the birthing bed
IMPORTANT The CPR function is in full agreement with the requirements of valid norms
To manually position the back rest:
Pull the red CPR leaver while
pushing the back rest of the bed
down. Keep holding the lever.
The CPR lever is located on the back rest
of the bed and it is accessible from either
side. The back rest of the bed sinks de-
pending on the weight of the patient
To reset the back section to its
original position:
Use motoric positioning (activates
via hand control, side control panel
or side rail panel)