Pressure
Symptom Solution
System pressure
too low
Check the following:
Pressure gauge calibration.
Ensure tank outlet valve is open.
Ensure pump inlet and outlet tubing are
connected correctly. Ensure Sta-Pure
(white) tubing is connected from pump
outlet to feed in port (FEED IN).
If theuidmovesbackandforthinthe
pump inlet or pump outlet tubing, par-
ticulates may be impairing the pump check
valves from closing. Ensure pump check
valvesareclean.Seecheckvalveushing
procedure in maintenance section “Pump
check Valve Flushing” of this manual
If ushingthecheckvalvesdoesnot
eliminate the problem, replace pump check
valves. See check valve replacement proce-
dure in maintenance section “Diaphragm
Pump Maintenance” of this manual
System Pressure
too high
Check the following:
Pressure gauge calibration
The retentate back-pressure valve is closed