Labscale Stirbase Setup
• If you order the pump module separatly, ensure pump module
wiring connections are correctly installed, see Chapter “Labscale
Pump Module set up”.
Module set up
Risk of electric shocks
• Unplug power cord prior to opening the blank panel from system base.
• There are no user-serviceable parts inside the Labscale system.
• Service should be performed only by trained and authorized
personnel.
Risk of rupture
• Use caution when opening and closing valves. Failure to open or
close valves at the appropriate times could result in trapped high
pressure, product loss, damaged cassettes, chemical spills, and
pump damage. Carefully follow all operating instructions
• Prior to each use, carefully inspect all hoses and tubing for kinks
or restrictions that could result in pressure buildup and possible
rupture of hoses or tubing.
• Prior to each use, carefully check that tubing at pump outlet to feed
in port (FEED IN) and cleaning tubing are Sta-Pure (white) tubing.
Failure in correctly installing Sta-Pure (white) tubing can result in
overpressurized tubing and burst.
• Transfer pump tubing can rupture due to continuous contact
with moving parts. Follow tubing manufacturer’s recommended
replacement interval.
• Never move the system while there is liquid in the tank, while
system processes are running or the cassettes is in place.