2-20
2.3.2 System Task
NO Task Function Period
1 System initialization Initializing the system
In case of a
startup
2 Data processing Analyzing and saving the data 1 second
3
Display of timer
information
Implementing the timed refreshing 1 second
5
Switchover of modules
and screens
Switching over between waveforms and
parameters on the screen
In case of a
screen change
event
6
Processing of user
commands and screens
Processing the user inputs by buttons and
displaying them on the screen.
In case of a
button event
7 System monitoring
System monitoring, voltage monitoring and
battery management
1 second
8 Network connection Implementing the network connection 1 second
9 Network data sending Sending the network data 1 second
10 Network data receiving Receiving the network data (viewbed) 1 second
11 ECG analysis
Analyzing ECG signal, calculating ECG values
(HR, ARR and ST), and saving the analysis
results.
1 second
12 Record output Outputting records
In case of a
record event
13 NIBP processing Implementing NIBP-related processing 1 second
14 WATCHDOG task Managing the system watchdog 1 second

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals