4-10
4.2.2 White Screen & Other Abnormal Screen
In case of faults causing white screen or other abnormal screens,
Check whether the LCD connection wires are in good contact;
Replace the LCD connection wires, or replace the LCD if necessary;
Replace the main control board if the fault still exists.
4.2.3 Encoder Faults
If all other functions (indicator, alarm, buttons) of the button panel are normal, proceed to
step 2; otherwise, replace the button panel;
Check whether short-circuit or abnormal open-circuit occurs in the encoder pad;
Replace the encoder.
4.2.4 No Audio Alarm
Check whether the audio alarm function is disabled in the software settings;
Replace the speaker;
Replace the button panel.
4.2.5 Printing Failure
Check whether there is any alarm about the recorder. If any, eliminate it;
Check whether the recorder indictor is on;
If not, check the connection wire for inputting signals to the recorder;
Check whether the recorder module is enabled in the maintenance menu;
Check the power cord of the recorder (including the recorder power PCB);
Replace the recorder module.
4.2.6 Abnormal Paper Drive
Check whether there are blocks on the paper roller of the recorder;
Check whether there are blocks in the gear cluster of thermal assembly of the recorder;
Check whether the voltage input of the recorder is larger than 17.6V.

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals