3-15
Nellcor SpO
2
Module
All SpO
2
sensors specified in the section Nellcor SpO2 Accessories meets the following
specifications when used with Nellcor SpO
2
module.
Sensor Range Precision*
MAX-A, MAX-AL, MAX-N,
MAX-P, MAX-I and MAX-FAST
70 to 100%
0% to 69%
±2%
Undefined
OxiCliq A, OxiCliq N, OxiCliq P,
OxiCliq I
70 to 100%
0% to 69%
±2.5%
Undefined
D-YS, DS-100A, OXI-A/N and
OXI-P/I
70 to 100%
0% to 69%
±3%
Undefined
SpO2 measurement range
and precision
MAX-R, D-YSE and D-YSPD
70 to 100%
0% to 69%
±3.5%
Undefined
PR measurement range and
precision
20 to 250 bpm: ±3 bpm
251 to 300 bpm: Undefined
Refreshing rate 1 s
Averaging time 8 s, 16 s
*: When sensors are used on neonatal subjects as recommended, the specified precision range is
increased by ±1%, to account for the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood.
3.11 NIBP Specifications
Measurement technique Auto oscillation
Displayed parameters Systolic pressure, diastolic pressure, mean pressure and PR
Mode of operation Manual, auto and continuous
Measurement interval in
auto mode
1/2/3/4/5/10/15/30/60/90/120/180/240/480 minutes
Measurement time in
continuous mode
5 minutes
mmHg Adult Pediatric Neonate
Systolic pressure 40 to 270 40 to 200 40 to 135
Diastolic pressure 10 to 210 10 to 150 10 to 100
Measurement range in
normal mode
Mean pressure 20 to 230 20 to 165 20 to 110

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals