3-4
3.5 Wireless network
Standards IEEE 802.11b, Wi-Fi compatible
Frequency range 2.412 to 2.462GHz
China America Canada Europe Spain France Japan
1 to 11 10, 11 2
Operating channel
For other country, please refer to your local law.
Safe distance 10m (a circle centering AP with the diameter of 10m)
Maximum data rate 11Mbps
3.6 Data Storage
Trend data
Long trend: 96 hours, resolution 1min, 5 min or 10 min.
Short trend: 1 hour, resolution 1 s or 5 s.
Alarm events
70 alarm events and associated waveforms (with user selectable
waveform length 8s, 16 or 32).
ARR events 80 ARR events and associated waveforms with 8s wavelength.
NIBP measurements
800 NIBP groups, including systolic pressures, mean pressures,
diastolic pressures and measurement time.
3.7 Signal Output Specifications
Standards
Meets the requirements of EC60601-1 for short-circuit
protection and leakage current
Output impedance 50Ω
ECG analog output
Bandwidth (-3dB; reference
frequency: 10Hz)
Diagnostic mode:
Monitor mode:
Surgery mode:
0.05 to 100 Hz (812A module)
0.05 to 150 Hz (M08A module)
0.5 to 40 Hz
1 to 20 Hz
Maximum propagation delay 25 ms (In DIAGNOSTIC mode, NOTCH is OFF)
Sensitivity 1 V/mV± 5%
PACE rejection/enhancement No pace rejection or enhancement