3-16
Measurement precision
Maximum average error: ±5mmHg
Maximum standard deviation: 8mmHg
Resolution 1mmHg
Static pressure
measurement range
0 to 300mmHg
Static accuracy ± 3 mmHg
Over-pressure protection
by software
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg
Over-pressure protection
by hardware
Adult:
Pediatric:
Neonate:
330 mmHg
330 mmHg
165 mmHg
Default start pressure
Adult:
Pediatric:
Neonate:
178±5 mmHg
133±10 mmHg
67±5 mmHg
PR from NIBP
Measurement range 40 to 240 bpm
Precision ±3 bpm or ±3%, whichever is greater
Resolution 1 bpm
3.12 TEMP Specifications
Number of channels 2
Displayed parameters T1, T2 and TD
Measurement range
0 to 50°C (32 to 122°F)
Resolution
0.1°C
Precision
±0.1°C (excluding the sensor)
±0.2°C (including the YSI 400 series sensor)
Update period 1s
Minimum time for
accurate measurement
Body surface: < 100s
Body cavity: < 80s
(YSI 400 series sensor)

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals