3-18
3.14 CO
2
Specifications
Measurement technique Infrared absorption technique
Displayed parameter EtCO2, FiCO2, Respiration Rate
CO2 function
Meet the requirements of EN ISO 21647/ISO 21647 and
ISO9918.
Mindray DS CO
2
Specifications
CO2 measurement range 0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2 mmHg
±5%
±10%
Resolution 1 mmHg
Drift meet the requirement of accurancy in 6 hours
Sample flow rate 70, 100 ml/min
Precision of deflation rate ±15% or 15 ml/min, whichever is great
Start-up time of CO2
module
< 1min, the module enters the warming up status after the startup.
One minute later, it enters the ready-to-measure status.
AwRR measurement range 0 to 120 BrPM
Precision
0 to 70 BrPM:
> 70 BrPM:
±2 BrPM
±5 BrPM
Response time
When measured with a neonatal watertrap and a 2.5 m-long
neonatal sampling line:
<3.5 s @ 100 ml/min
<4 s @ 70 ml/min
When measured with an adult watertrap and a 2.5 m-long adult
sampling line:
<5.5 s @ 100 ml/min
<7 s @ 70 ml/min
Delay time
When measured with a neonatal watertrap and a 2.5m-long
neonatal sampling line:
<3 s @ 100 ml/min
<3.5 s @ 70 ml/min
When measured with an adult watertrap and a 2.5m-long adult
sampling line:

Table of Contents

Other manuals for Mindray DPM 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DPM 4 and is the answer not in the manual?

Mindray DPM 4 Specifications

General IconGeneral
BrandMindray
ModelDPM 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals