Services Calibration Procedure
5 - 4 0070-10-0705 Passport V™ Service Manual
5.4.2 NIBP Maintenance
The NIBP Maintenance offers the choice of NIBP Accuracy Test, NIBP Leakage Test, and
Calibrate NIBP.
FIGURE 5-3 NIBP Maintenance Menu
5.4.2.1 NIBP Accuracy Test
1. Connect the equipment as shown in FIGURE 5-4.
2. Select NIBP Accuracy Test using the Navigator
TM
Knob to activate the test.
3. Using the manometer’s bulb valve, raise the pressure in the test chamber to 300mmHg.
4. Compare the pressure displayed on the screen with the pressure viewed on the
manometer. The difference between readings should not be greater than +/-3 mmHg.
NOTE: An NIBP simulator can be used to replace the reference
manometer with the bulb valve to perform the test.
Specifications: 0 to 300 mmHg +/- 3 mmHg