Sterilization and Cleaning of Cuffs Preventative Maintenance
6 - 8 0070-10-0705 Passport V™ Service Manual
6.6.2 Reusable Bladderless Cuffs
Clean cuffs with warm water and a mild detergent. Do not use a detergent containing hand
conditioners, softeners, or fragrances.
NIBP cuffs can be sterilized with gamma sterilization without affecting the repeated
performance of the cuff. Steam sterilization is not recommended. Use of a washing liquid
containing bleach is not recommended because chlorine will chemically break down the
urethane on the inside of the cuff.
Antimicrobial Definition
Bladderless cuffs are treated with an antimicrobial coating. Antimicrobial technology
effectively controls a broad spectrum of bacteria, fungi, algae and yeasts on a wide variety
of treated substrates.
6.6.3 Disposable Blood Pressure Cuffs
Disposable cuffs are intended for single patient use only. Once a cuff is used on a patient it
should be discarded. Do not use the same cuff on any other patient. Do not sterilize or use an
autoclave on disposable cuffs.
CAUTION: Disposable cuffs can be cleaned using a mild soap solution
and dried with a clean cloth.

Table of Contents

Other manuals for Mindray Passport V

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish