EasyManua.ls Logo

Mindray Passport V Service Manual

Mindray Passport V
131 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Troubleshooting Guidelines Repair Information
3 - 2 0070-10-0705 Passport V™ Service Manual
3.3 Troubleshooting Guidelines
1. Identify the problem
Due to the wide variety of potential symptoms, certain problems may be more subtle
than others. One approach to troubleshooting is to set up the instrument as described in
Chapter 5.0. Following the guidelines of the tests will help determine the problem if one
exists.
2. Avoid shorting component leads
During repair procedures, it can become tempting to make a series of quick
measurements. Always turn the power off before connecting and disconnecting the test
leads and probes. The accidental shorting of leads can easily stress the components and
cause a second failure (aside from the safety risk).
3. Use the Proper equipment
The equipment listed below is suggested to fulfill a wide range of troubleshooting
requirements. It is imperative to use the designated equipment in order to ensure proper
results of any and all test procedures.
4. Clean up the repair area
After any repair, especially after any soldering or desoldering, clean off the repair area
with alcohol and a stiff brush. This will remove any residual solder flux, in turn allowing
the instrument to return to its original appearance.
3.4 Special Tools Required
•Digital Voltmeter
•Digital Mercury Manometer 0 to 300 mmHg
•Safety Analyzer
•Patient Simulator
•Test Chamber / Dummy Cuff (P/N 0138-00-0001-01 (700 cc) or -03 (500 cc))
•Desktop PC or notebook PC.
•Microsoft Windows 2000/XP operating system
•Intel Pentium CPU, above 500MHz
•Above 128M memory
•At least one network card and at least one USB port
•CAT-5 crossover cable
USB cable or DPM SB storage device
3.5 Disassembly Instructions
Before disassembling the unit, perform the following:
•Power down the unit and remove the line cord.
•Remove all cable assemblies from the left side, right side, and rear of the unit.
•Remove any batteries that were installed.
•Perform all work on a properly grounded ESD workstation.

Table of Contents

Other manuals for Mindray Passport V

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport V and is the answer not in the manual?

Mindray Passport V Specifications

General IconGeneral
BrandMindray
ModelPassport V
CategoryHealth Care Products
LanguageEnglish

Summary

Foreword

Warnings, Cautions, and Notes

General guidelines and important safety information for users and technicians.

Warnings

Alerts regarding potential serious risks, including electrical shock hazards.

Cautions

Notes on using the device with care to avoid adverse effects or damage.

Theory of Operation

Repair Information

Safety Precautions

Essential safety guidelines and warnings for personnel performing repairs.

Troubleshooting Guidelines

Methods and advice for diagnosing and resolving instrument problems effectively.

Disassembly Instructions Overview

General steps and precautions before disassembling the patient monitor unit.

Troubleshooting Menus

ECG Troubleshooting

Common ECG messages, reasons, and solutions for troubleshooting.

NIBP Troubleshooting

Common NIBP messages, reasons, and solutions for troubleshooting.

SpO2 Troubleshooting

Common SpO2 messages, reasons, and solutions for troubleshooting.

Temperature Troubleshooting

Common temperature messages, reasons, and solutions for troubleshooting.

Resp Troubleshooting

Common respiration messages, reasons, and solutions for troubleshooting.

IBP Troubleshooting

Common IBP messages, reasons, and solutions for troubleshooting.

CO2 Troubleshooting

Common CO2 messages, reasons, and solutions for troubleshooting.

Gas Module Troubleshooting

Common gas module messages, reasons, and solutions for troubleshooting.

Installation Menu

Installation Mode

Procedure for entering and configuring the monitor's installation settings.

Advanced Installation Setup Menu

Configuration options for network, IP address, and other advanced settings.

Calibration Procedure

Warnings and Guidelines

Safety precautions and general guidelines for performing calibration procedures.

Verification Procedures

Step-by-step guides for verifying the function of ECG, NIBP, SpO2, etc.

Preventative Maintenance

Preventative Maintenance Schedule Overview

List of recommended periodic maintenance activities and intervals.

Electrical Safety Tests

Procedures for performing electrical safety and leakage current tests.