4-7
4.4.3 CO
2
Test
Accuracy test
Tool required:
A steel gas cylinder with 5±0.03% CO
2 and balance gas N2 (P/N 0075-00-0033-01)
T-shape connector
Tubing
1. Select Main] [Maintenance >>][User Maintenance>>]enter the required
password [Module Maintenance>>][Calibrate CO2>>]
2. Connect the test system as follows:
3. Open the valve to flow CO
2 and make sure that there is flow sufficient to vent to
atmosphere.
4. Verify the realtime CO
2 value is within 5.0 ± 0.3% in the [Calibrate CO2] menu.
CO2 calibration
Tool required:
A steel gas cylinder with 5±0.03% CO
2 and balance gas N2 (P/N 0075-00-0033-01)
T-shape connector
Tubing
1. Select [Main] [Maintenance >>][User Maintenance>>]enter the required
password [Module Maintenance>>][Calibrate CO2>>].
2. In the [Calibrate CO2] menu, select [Zero].
3. After the zero calibration is finished successfully, connect the equipment as follows:
Flowmeter
Sampling line
T-shape connector
Cylinder
Relief value
Monitor

Table of Contents

Other manuals for Mindray VS-900

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray VS-900 and is the answer not in the manual?

Mindray VS-900 Specifications

General IconGeneral
BrandMindray
ModelVS-900
CategoryMonitor
LanguageEnglish

Related product manuals