6-10
6.3.9 Disassembling the CO
2
Module Assembly
1. Pull out the AC receptacle, NIBP connection cable, CO2 connection cable.
2. Unscrew two M3X6 screws.
3. Remove the CO
2
module.
6.3.10 Disassembling Pumps and Valves (NIBP optional)
1. Cut the two cable ties and take out the gas pump.
CO
2
connection
cable
Screw
NIBP connection
cable
AC receptacle
Screw
Cable tie