Chapter 3 Assembly
64
Chapter 3
Assembly
6
Assembly for Epi-fluorescence Microscopy
Follow the procedure described in Section 2 “Assembly for Bright-field Microscopy” to perform assembly.
Note the following:
Attaching the epi-fluorescence attachment
Attach the epi-fluorescence attachment to the microscope arm prior to the step 4, “Attaching the tube.”
(1) Place the epi-fluorescence attachment on the
microscope arm.
(2) Secure it by tightening the clamp screw on the arm and
two fixing bolts at the rear of the epi-fluorescence
attachment. Use a hex wrench provided with the
epi-fluorescence attachment to tighten the fixing bolts.
(3) Attach the HG adapter of the mercury lamp illuminato
r
to be used to the bayonet mount on the back. (Fo
r
detailed information, refer to the instruction manual
provided with the mercury lamp illuminator.)
(4) Place the tube on the epi-fluorescence attachment and
tighten the tube clamp screws of the epi-fluorescence
attachment using the tool provided with the microscope
to secure the tube.
A1
-
2
-
3
-
4
B
1
-
2 / 3
-
4
C
1
-
2
-
3
-
4
1
1
4
2
3
4
CUBE
Attaching the epi-fluorescence attachment
Attaching a filter cube
WARNING
Always turn off the mercury lamp before inserting or removing a filter cube.
(1) Remove the filter cube replacement cover on the left o
f
the epi-fluorescence attachment.
(2) Insert a filter cube.
The filter cubes listed below cannot be inserted directly
into the filter bay of the epi-fluorescence attachment.
Remove an internal spacer or reverse the space
r
before insertion.
<Filter cubes requiring spacer removal>
UV-2A
UV-2B
<Filter cubes requiring spacer reversal>
DAPI
FITC
GFP-L
GFP-B
TRITC
Tx-Red
A1
-
2
-
3
-
4
B
1
-
2 / 3
-
4
C
1
-
2
-
3
-
4
A
UV-2A
EX 330-380
DM 400
BA 420
ND4
ND8
ND16
A1
-
2
-
3
-
4
B1
-
2 / 3
-
4
C1
-
2
-
3
-
4
Attaching a filter cube
Epi-fluorescence
attachment
clam
p
screw
Epi-fluorescence
attachment
HG adapte
r
Epi-fl attachment clamp bolt
(use a tool)
Tube clamp
screw
(use a tool)
Filter cube
replacement cover
Filter cube

Table of Contents

Other manuals for Nikon eclipse Ci-L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon eclipse Ci-L and is the answer not in the manual?

Nikon eclipse Ci-L Specifications

General IconGeneral
BrandNikon
Modeleclipse Ci-L
CategoryMicroscope
LanguageEnglish

Related product manuals