Chapter 4 Troubleshooting
72
Chapter 4
Troubleshooting
1.3
Phase Contrast Microscopy
Problem Cause Measure
The Ph annular diaphragm of the
condenser does not match the phase plate
image of the objective.
Adjust so that they match.
(Chapter 1, “2.2 Phase Contrast Microscopy
Procedure 12 Center the Ph annular diaphragm”)
The Ph annular diaphragm of the
condenser and the objective selected do
not match.
Put the Ph annular diaphragm with the same
Ph code as the objective into the optical path.
(Chapter 1, “2.2 Phase Contrast Microscopy
Procedure 14 Adjust the Ph annular diaphragm in
the condenser with the Ph objective to be used.”)
The phase contrast of the specimen is too
large.
Change the mounting agent or thickness of
the specimen when preparing the specimen.
(Chapter 2, “14 Tips for Phase Contrast
Microscopy”)
Poor contrast.
The type of the Ph objective is not suitable
for phase contrast of the specimen.
Use a Ph objective suitable for the specimen.
(Chapter 2, “14 Tips for Phase Contrast
Microscopy”)

Table of Contents

Other manuals for Nikon eclipse Ci-L

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nikon eclipse Ci-L and is the answer not in the manual?

Nikon eclipse Ci-L Specifications

General IconGeneral
BrandNikon
Modeleclipse Ci-L
CategoryMicroscope
LanguageEnglish

Related product manuals