51
Upper Limit (H) of Linear Output Range
Lower Limit (L) of Linear Output Range
Linear output setting is made in the option menu to enable the K3NX to have
voltage or current output in proportion to the change in display value.
The maximum and minimum values of linear output are set in this parameter.
Linear output
20 mA or 5 V
Display value
4 mA or 1 V
L can be greater or less than H.
L cannot be the same as H, otherwise H will be automatically set to a value
obtained by adding 1 to L.
The teaching function can be used for setting linear output ranges.
Input type Setting range Default
DC voltage input 19999 to 99999
H linear output range
199.99
L linear output range 199.99
DC current input 19999 to 99999
H linear output range 199.99
L linear output range 199.99
AC voltage input 19999 to 99999
H linear output range 400.0
L linear output range 0.0
AC current input 19999 to 99999
H linear output range 10.000
L linear output range 0.000
Refer to 5-1 Teaching Function.
This setting is available for the K3NX with the Linear Output Board.
lset.h
lset.l
FUNCTION
SETTING
REFERENCE
MODELS
Setting Mode
Section 3-2
AUDIN - 7 bis rue de Tinqueux - 51100 Reims - France - Tel : 03.26.04.20.21 - Fax : 03.26.04.28.20 - Web : http: www.audin.fr - Email : info@audin.fr

Table of Contents

Other manuals for Omron K3NX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron K3NX and is the answer not in the manual?

Omron K3NX Specifications

General IconGeneral
BrandOmron
ModelK3NX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals