Page #88 background imageLoading...
Page #88 background image
80
5-3 Maintenance Mode
The set values of the K3NX can be initialized and field calibration of the K3NX is
possible in maintenance mode. The operations of the K3NX in this mode are de-
scribed below.
5-3-1 Maintenance Mode
The K3NX will be in maintenance mode if the Mode and Shift Keys are pressed
simultaneously while the K3NX is turned on.
The K3NX in maintenance mode will go into RUN mode if the Escape Key is
pressed.
RUN mode
Press the Mode and Shift Keys
while turning the K3NX on.
Power On
Press the Escape and Up
Keys for 1 second.
Maintenance mode
RUN mode
1 second
Setting mode
Protect mode
5-3-2 Initialization
Follow the steps described below to reset the set values of the K3NX to factory-
set values.
1, 2, 3... 1. Press the Mode and Shift Keys simultaneously while turning the K3NX on.
The mant maintenance mode setting will be displayed.
Set Value LED Display Model Basic Model
2. Press the Mode Key for more than one second. The reĆin initialization set-
ting will be displayed.
Set Value LED Display Model Basic Model
3. Press the Up Key to display no.
Set Value LED Display Model Basic Model
Maintenance Mode
Section 5-3
AUDIN - 7 bis rue de Tinqueux - 51100 Reims - France - Tel : 03.26.04.20.21 - Fax : 03.26.04.28.20 - Web : http: www.audin.fr - Email : info@audin.fr

Table of Contents

Other manuals for Omron K3NX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron K3NX and is the answer not in the manual?

Omron K3NX Specifications

General IconGeneral
BrandOmron
ModelK3NX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals