Network operation and remote control
R&S
®
LCX Series
153User Manual 1179.2260.02 ─ 02
Remote command:
SYSTem:COMMunicate:LAN:ADDRess on page 217
Subnet Mask
Displays the bit group of the subnet in the host identifier.
To assign the subnet mask manually, set DHCP & Auto-IP > Off.
How to: see "Assigning the LAN interface addresses manually" on page 159.
Remote command:
SYSTem:COMMunicate:LAN:SMASk on page 218
Gateway
Displays the gateway address.
This address identifies the router on the same network as the instrument that is used to
forward traffic to destinations beyond the local network.
To assign the subnet mask manually, set DHCP & Auto-IP > Off.
How to: see "Assigning the LAN interface addresses manually" on page 159.
Remote command:
SYSTem:COMMunicate:LAN:DGATeway on page 217
Apply Configuration
Applies the modified settings.
Remote command:
n.a.
Reset LAN Conection
Terminates the network connection of the instrument and sets it up again.
You can use this function to fix network problems.
Note: This function restarts only the connection of the instrument to the network. It
does not impact the network itself.
How to: see "Assigning the LAN interface addresses manually" on page 159.
Remote command:
SYSTem:COMMunicate:LAN:RESet on page 218
MAC Address
Indicates the MAC address, a unique identifier of the network adapter in the R&S LCX.
Remote command:
SYSTem:COMMunicate:LAN:MAC? on page 218
10.3.3 VNC settings
Using a VNC client tool on a computer, you can control the R&S LCX remotely. Estab-
lish a connection between the computer and R&S LCX in the network, see "Establish-
ing the LAN connection" on page 21.