Instrument functions
R&S
®
LCX Series
73User Manual 1179.2260.02 ─ 02
[Auto].............................................................................................................................73
[Hold].............................................................................................................................73
[Trig.]............................................................................................................................. 73
[Auto]
Selects the impedance range automatically, according to the connected sample com-
ponent.
Note: If you have fixed the impedance range with [Hold] > ON, automatic selection is
blocked.
Remote command:
FUNCtion:IMPedance:RANGe:AUTO on page 196
[Hold]
Freezes the currently set range.
When activated, the R&S LCX displays the prefix "Hold:" at the range value.
Remote command:
FUNCtion:IMPedance:RANGe:HOLD on page 196
[Trig.]
Triggers one measurement cycle manually.
Remote command:
*TRG on page 175
8.6 Measurement parameters
The settings described in this section refer to parameters that you can configure for
your specific measurement in addition to the test signal, measurement mode, or mea-
surement functions.
About measurement parameters
Aspects on the measurement parameters:
Measurement speed
The measurement speed defines the number of measurements in one second
when triggered continuously.
Impact of measurement speed and accuracy:
The measurement accuracy decreases with increasing speed, i.e. the higher the
number of measurements per second you perform, the lower the measurement
accuracy you can expect.
At low test signal frequencies, the R&S LCX reduces the measurement speed
automatically.
See "Measurement Speed" on page 76.
Cable length
Any cable covers inductive and capacitive parts, which depend in particular on the
cable length. As these characteristics are considered as parasitics, they distort a
Measurement parameters

Table of Contents

Other manuals for R&S LCX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S LCX Series and is the answer not in the manual?

R&S LCX Series Specifications

General IconGeneral
BrandR&S
ModelLCX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals