Measurement setups
R&S
®
LCX Series
56User Manual 1179.2260.02 ─ 02
Figure 6-5: R&S
LCX-Z5 transformer test fixture and test cables
The test fixture comes with four test leads equipped with alligator clips, to contact the
primary and secondary circuits of the measurement sample. The BNC connectors on
the right and left outside are intended to contact the circuit with the higher number of
windings. The two BNC connectors in the middle correspondingly for the circuit with
lower number of windings.
As an alternative, it is possible, to install a user-specific test setup at the BNC connec-
tors directly.
For open and short correction, the test leads must be connected to the test fixture as
follows:
1. Connect the test leads with the red alligator clips to the "N", and to the outer "Cus-
tom" [BNC] connectors.
2. Connect the test leads with the black clips to the "I" and the inner "Custom" [BNC]
connectors.
To perform short correction, connect the two red clips and the two black clips together.
For open correction, connect the black clips together and leave the red clips open.
About test fixtures

Table of Contents

Other manuals for R&S LCX Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S LCX Series and is the answer not in the manual?

R&S LCX Series Specifications

General IconGeneral
BrandR&S
ModelLCX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals