Technical Data
MP150 Rev. E3 07/2013 13
2.3 Electrical Specifications
Communications
Ethernet TCP/IP protocol 10/100 MBit/s, electrically isolated,
auto-negotiation
RS485 full-duplex, not addressable
9.6, 57.6, 115.2, 230.4 kBaud
Outputs
Analog 3 active current outputs, each adjustable 0/4 ... 20 mA,
maximum load: 500 Ω, minimal load: 47 Ω
common ground connection for all current outputs, electrically
isolated to the GND ground
Alarm potential-free relay contacts: 30 V / 1 A
normally open / normally closed
Inputs
Trigger + 5 to 24 VDC pulse, high/low active
Functional Input max. + 5 VDC
function depends on the specific MP150 based system
Signal Processing
Stand-alone unit Max, Min, Average, Peak/Valley Hold, Alarm setpoints
further signal processing function through software
Power Supply 24 VDC ± 25%, 1 A
2.4 General Specifications
Mechanical Scanning System 40,000 h (about 4.5 years) meantime between failures (MTBF)
for the mechanical scanning system at 48 Hz scan rate
Environment Rating IP65 (NEMA 4) in accordance to IEC 60529
Ambient Temperatures 0 to 50°C (32 to 122°F) – without water cooling
max. 180°C (356°F) – with water cooling
down to –40°C (–40°F) – only for scanner with internal heating
element (optional) and still air (e.g. by using a protective
housing)
Storage Temperatures -25 to 65°C (-13 to 149°F)
Water Cooling max. pressure 15 bar (218 PSI)
Air Purge Collar max. pressure 3 bar (43 psig)
Warm-up Time 30 min.
Relative Humidity 10 to 90%, non-condensing,
for operating and non-operating temperature range
Vibration IEC 60068-2-6, 3 axes, 10 to 150 Hz, operating 2 g above 20 Hz
Shock IEC 60068-2-27, 3 axes, operating: 5 g at 11 ms, 15 g at 6 ms
Dimensions 200 x 180 x 190 mm (7.8 x 7.02 x 7.41 in)