13.1.1 Working mode selection
Parameter allowing to select method of determination of sample piece mass.
1. Enter <3.2.PcS / 3.2.2.UUt> submenu.
2. Go to <Acc> function.
3. Press key, parameter values are displayed successively one by one, where: S_S – Select
to set sample mass by determining mass of a single part, Suu – Select to set sample mass by
entering mass of a single part.
4. Enter respective value and press key to confirm, then continue weighing.
13.2 Setting reference sample mass by entering mass of a single part
1. Enter <3.2.PcS / 3.2.2.UUt> submenu, set <Suu> value.
2. Enter <PcS> working mode (Parts Counting), first, text <Set_Ut> is displayed for 1s, next you
see a window for entering mass value of a single part.
3. Enter respective value and press key to confirm, the home screen is displayed
automatically along with the quantity of parts loaded onto the weighing pan (pcs).
Note: If the value of entered single part mass is greater than max capacity value, then message <Err
Hi> is displayed.
13.3 Setting reference sample mass by determining mass of a single part
1. Enter <3.2.PcS / 3.2.2.UUt> submenu, set <S_S> value.
2. Enter <PcS> working mode (Parts Counting), first, text <Set_Ut> is displayed for 1s, next you
see a window for entering mass value of a single part.
3. Press key to select one of the following options: 10 – 10 pieces, 20 – 20 pieces, 50 – 50
pieces, 100 – 100 pieces, 0000 – Custom reference sample quantity to enter required value.
4. Select respective option and press key to confirm, first, text <LoAd> is displayed for 1
second, next you see a weighing window.
5. If the parts are to be weighed in a container, first put the container on a weighing pan and
tare it.
6. Load the weighing pan with the declared amount of parts. When the indication is stable (
pictogram is displayed), press key to confirm the mass.
7. Single part mass is calculated automatically, the quantity of parts (pcs) is displayed.
Note: Total weight value of all parts loaded onto the weighing pan cannot be greater than the max
capacity value.
Note: Single part mass value must be equal or greater than 0.1 of the reading unit. Unless this
condition is met, the weighing device displays a message <ErrLo>.
Note: In the course of parts quantity determination it is necessary to wait for a stable measurement
before confirming the declared quantity value.