RIGOL Chapter 1 Specifications
Service Guide for DS1000E, DS1000D Series
1-2
Specifications
Acquisition
Sampling Modes
Real-Time
Equivalent
Sampling Rate
1GSa/s
[1]
, 500MSa/s
DS1102X
DS1052X
25GSa/s
10GSa/s
Averages
N time acquisitions, all channels simultaneously, N is
selectable from 2, 4, 8, 16, 32, 64, 128 and 256.
Inputs
Input Coupling
DC, AC, GND
Input Impedance
1MΩ±2%
18pF±3pF
Probe Attenuation Factors
1X, 5X, 10X, 50X, 100X, 500X,1000X
Maximum Input Voltage
400V (DC
AC Peak, 1MΩ input impedance)
40V (DC+AC Peak)
[1]
Time delay between channel
(typical)
500ps
Horizontal
Sample Rate
Range
Real-Time: 13.65Sa/s-1GSa/s
Equivalent: 13.65Sa/s-25GSa/s
Waveform
interpolation
Sin(x)/x
Record Length
Channel Mode
Sample rate
Record
Length
(normal)
Record
Length
(long record)
Single channel
1GSa/s
16kpts
N.A.
Single channel
500MSa/s
Or lower
16kpts
1Mpts
Double channel
500MSa/s
Or lower
8kpts
N.A.
Double channel
250MSa/s
Or lower
8kpts
512kpts
Scan speed Range
(Sec/div)
2ns/div~50s/div, DS1102X
5ns/div~50s/div, DS1052X
1-2-5 Sequence
Sample Rate and
Delay Time
Accuracy
±50ppm (over any 1ms time interval)
Delta Time
Measurement
Accuracy
(Full Bandwidth)
Single-shot: ±(1 sample interval + 50ppm × reading + 0.6 ns)
>16 averages: ±(1sample interval + 50ppm × reading + 0.4
ns)
Vertical

Table of Contents

Other manuals for Rigol DS1102D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1102D and is the answer not in the manual?

Rigol DS1102D Specifications

General IconGeneral
BrandRigol
ModelDS1102D
CategoryTest Equipment
LanguageEnglish

Related product manuals