RIGOL Chapter 3 Performance Test
Service Guide for DS1000E, DS1000D Series
3-8
P/F Interface Test
Purpose:
Test if P/F interface work normally.
Tools:
A set of DS1000E, DS1000D series digital oscilloscope
A BNC cable
Steps:
1. Power on the DS1000E and DS1000D respectively.
2. Connect the P/F interface at the rear panel of the DS1000E and the CH1 channel
of the DS1000D using the BNC cable.
Figure 3-10 P/F interface connection
3. Turn on the oscilloscope, press Storage Storage Factory to recall the
factory settings.
4. Press UtilityPass/fail to enter the setting interface of Pass/Fail, see the table
below.
Table 3-1 Pass/Fail setting items
Items
Setting
Enable Test
ON
Source
CH1
Msg display
ON
Output
Fail
Stop on Output
OFF
5. After setting, press Operate to start Pass/Fail test.
6. When the test failed, observe the CH1 channel of the DS1000D. If a pulse
waveform is displayed, it indicates that the interface works normally. If there is no
response, errors might be happened to the interface. At this point, you need to
inspect or mend the interface.
7. Using the same method to test if P/F interface of the DS1000D work normally.

Table of Contents

Other manuals for Rigol DS1102D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1102D and is the answer not in the manual?

Rigol DS1102D Specifications

General IconGeneral
BrandRigol
ModelDS1102D
CategoryTest Equipment
LanguageEnglish

Related product manuals