Chapter 3 Performance Test RIGOL
Service Guide for DS1000E, DS1000D Series
3-11
Specifications Test
This section introduces the performance verification test method of
DS1000E/DS1000D series digital oscilloscope by taking DS1102E as an example. You
can perform the following test in any order. In this manual, the test device used is
Fluke 9500B. You can also use other devices that fulfill the specification requirements
for the test.
Recommended Devices List:
Device Specification
Recommended
Model
Oscilloscope
Calibrator
Output range of DC voltage:
1 MΩ: 1 mV to 200 V
50 MΩ1 mV to 200 V
The rise time of fast edge signal: ≤ 150
ps
Fluke 9500B
Digital
Multimeter
The resistance measurement accuracy is
higher than ±0.1% of reading
RIGOL DM3058/3068
Test Cable
BNC (male) to Dual-banana Plug Cable
--
Signal Generator Frequency Accuracy: ±1 ppm RIGOL DG4162
Test Cable
BNC (m)-BNC (m) cable
--
Note:
1) Make sure that the oscilloscope passes the self-calibration before performing the
performance calibration test.
2) Make sure that the oscilloscope has been warmed up for 30 minutes before
executing any of the following tests.
3) Please reset the instrument to the factory setting before or after executing any of
the following tests.

Table of Contents

Other manuals for Rigol DS1102D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1102D and is the answer not in the manual?

Rigol DS1102D Specifications

General IconGeneral
BrandRigol
ModelDS1102D
CategoryTest Equipment
LanguageEnglish

Related product manuals