RIGOL Chapter 3 Performance Test
Service Guide for DS1000E, DS1000D Series
3-18
17) Repeat step 6).
18) Output a sine waveform with 20 MHz frequency and 1.2 Vpp amplitude from
Fluke 9500B.
19) Repeat step 8), 9) and 10).
20) Output a sine waveform with 50 MHz frequency and 1.2 Vpp amplitude from
Fluke 9500B.
21) Repeat step 12), 13) and 14).
22) Keep other settings of the oscilloscope unchanged and set the vertical scale
to 500 mV/div.
23) Output a sine waveform with 1 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
24) Repeat step 6).
25) Output a sine waveform with 20 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
26) Repeat step 8), 9) and 10).
27) Output a sine waveform with 50 MHz frequency and 3 Vpp amplitude from
Fluke 9500B.
28) Repeat step 12), 13) and 14).
29) Turn CH1 off. Test CH2 according to the above test steps and record the test
results.