A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
D-1
Appendix D Maximum Acoustic Output Report
Table1-1. Transducer Model: 6-2MHz C351 Curved Array
Operating Mode: B-mode
TIS TIB
Non-scan
Index label MI
Scan
A
aprt
1cm
2
A
aprt
>1cm
2
Non-
Scan
TIC
Maximum index value
0.66 0.09 # # # #
p
r, α
(MPa)
1.15
P (mW)
5.9 # # #
min of (mW)
#
z
s
(cm)
#
z
bp
(cm)
#
z
b
(cm)
4.5
#
z at max I
pi.α
(cm)
4.5
#
d
eq
(z
b
) (cm)
#
f
awf
(MHz)
3.1 3.1 # # # #
Dim of A
aprt
X (cm)
1.9 # # # #
Associated
acoustic
parameters
Y (cm)
1.3 # # # #
t
d
(µsec)
0.56
prr (Hz)
4219
p
r
at max I
pi
(MPa)
1.84
d
eq
at max I
pi
(cm)
#
I
pi.α
at max MI (W/cm
2
)
32.0
FL
x
(cm)
5 # # #
Other
information
Focal
Length
FL
y
(cm)
5 # # #
Mode
B B # # # #
Focus
(cm)
5 5 # # # #
PRF
(kHz)
4.2 4.2 # # # #
Operating
control
conditions
Power
(%)
100 100 # # # #