EasyManuals Logo
Home>Sonoscape>Medical Equipment>A6

Sonoscape A6 User Manual

Sonoscape A6
172 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #88 background imageLoading...
Page #88 background image
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-66
3.6.2.3.6 R. Follicle (Right Follicle) measurement
Function: To measure length and width of right follicle: R.FOL, R.FOW.
Measurement method:
1. Press the Calckey to pop up the Meas &
Calc menu, move the trackball, use the cursor
to select “GYN”, press theSETkey to go to
the submenu, move the trackball to select R.
Follicle, press theSETkey to go to R.
Follicle measurement mode.
2. Measure R.FO L of right follicle; see “Distance
Measurement” for the specific method.
3. Measure R.FO L of right follicle; see “Distance
Measurement” for the specific method.
4. The measurement result is displayed in the
result area.
5. Repeat steps (1) to (4) to go to the next R.
Follicle measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sonoscape A6 and is the answer not in the manual?

Sonoscape A6 Specifications

General IconGeneral
BrandSonoscape
ModelA6
CategoryMedical Equipment
LanguageEnglish

Related product manuals