A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-61
3.6.2.3.1 UT measurement
Function: Measure length, height and width of uterus in sequence
UT-L Uterus length (mm)
UT-H Uterus height (mm)
UT-W Uterus width (mm)
UT = UT-L + UT-H + UT-W
Measurement method:
↓
↓
1. In B mode, scan and freeze the image, press
the【Calc】key to pop up the Meas & Calc
menu, move the trackball, use the cursor to
select “GYN”, press the【SET】key to go to the
submenu, move the trackball to select “UT”,
press the【SET】key to go to “UT”
measurement mode.
2. Measure UT-L, see “Distance Measurement”
for the specific method.
3. Measure UT-H, see “Distance Measurement”
for the specific method.
4. Measure UT-W, see “Distance Measurement”
for the specific method
5. UT = UT-L + UT-H + UT-W, the measurement
& calculation result is displayed in the result
window.
6. Repeat steps 1 to 5 to go to the next UT
measurement.