EasyManuals Logo
Home>Tektronix>Test Equipment>2205

Tektronix 2205 Service Manual

Tektronix 2205
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
Adjustment Procedure2205 Service
g. Return the CH 1 VOLTS/DIV Variable control to
the CAL detent
3. Adjust Channel 2 Invert Balance (R84)
a. Position the trace on the center horizontal
graticule line using the Channel 2 POSITION control.
b. Set Vertical MODE switch to CH 2 INVERT.
c. ADJUSTCH 2 Invert Bal (R84) to set the trace
to the center horizontal graticule line.
d. Set Vertical MODE switch to NORM.
rotate the CH 1 VOLTS/DIV Variable control fully
counterclockwise and CHECK that the display
decreases to two divisions or less. Then return the
CH 1 VOLTS/DIV Variable control to the CAL detent
and continue with the 50-mV check.
b. Move the cable from the CH 1 OR X input con
nector to the CH 2 OR Y input connector. Set the
Vertical MODE switch to CH 2.
c. Repeat part b using the Channel 2 controls.
Table 5-3
Deflection Accuracy Limits
e. Repeat parts a through d until there is no trace
shift when switching from NORM to CH 2 INVERT.
4. Adjust Vertical Gain (R145 and R195)
a. Set:
Vertical MODE
CH 1 VOLTS/DIV Variable
AC-GND-DC (both)
Trigger SOURCE
CH 1, NORM
CAL detent
DC
VERT MODE
b. Connect a 20-mV, standard-amplitude signal
from the calibration generator via a 50-D coaxial
cable to the CH 1 or X input connector.
c. Center the display within the graticule using
the CH 1 POSITION control.
d. ADJUST-CH 1 Gain (R145) for an exact 4-
division display.
e. Move the test-signal cable from the CHI or X
input connector to the CH 2 OR Y input connector.
f. Set the Vertical MODE switch to CH 2.
g. Center the display within the graticule using
the CH 2 POSITION control.
h. ADJUST-CH 2 Gain (R195) for an exact 4-
division display.
VOLTS/DIV
STANDARD
ACCURACY
Switch
Amplitude
Limits
Setting
Signal (Divisions)
5 mV
20 mV 3.88 to 4.12
10 mV
50 mV
4.85 to 5.15
20 mV
0.1 V 4.85 to 5.15
50 mV
0.2 V 3.88 to 4.12
0.1 V
0.5 V
4.85 to 5.15
0.2 V
1 V
4.85 to 5.15
...
0.5 V
..
2 V
3.88 to 4.12
1 V
5 V
4.85 to 5.15
2 V 10 V 4.85 to 5.15
5 V 20 V 3,88 to 4.12
6. Check Input Coupling
a. Set the AC-GND-DC switches (both channels)
to GND.
b. Position the trace on the center horizontal
graticule line using the CH 2 POSITION control.
c. Change the generator output to 50 mV.
d. Set the Channel 2 AC-GND-DC switch to AC.
i. Repeat parts b through h until the gain of the
two channels is identical.
e. CHECKThe display is centered about the
center horizontal graticule line.
5. Check Deflection Accuracy and VOLTS/DIV
Variable Range
a. CHECKDeflection accuracy is within the limits
given in Table 5-3 for each CH 1 VOLTS/DIV switch
setting and corresponding standard-amplitude sig
nal. When at the 20-mV VOLTS/DIV switch setting,
f. Set the Channel 2 AC-GND-DC switch to DC.
g. CHECKThe display is ground referenced on
the center horizontal graticule line.
h. Move the test-signal cable from the CH 2 OR Y
input connector to the CH 1 OR X input connector,
5-7

Table of Contents

Other manuals for Tektronix 2205

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2205 and is the answer not in the manual?

Tektronix 2205 Specifications

General IconGeneral
BrandTektronix
Model2205
CategoryTest Equipment
LanguageEnglish

Related product manuals