Performance Check Procedure—2230 Service
Table 4-1
Test Equipment Required
Item and
Description
Minimum
Specification
Purpose
Example of Suitable
Test Equipment
1. Calibration
Generator
Standard-amplitude signal levels:
5 mV to 50 V. Accuracy: ± 0.3%.
High-amplitude signal levels: 1 V to
60 V. Repetition rate: 1 kHz.
Fast-rise signal level: 1 V. Repetition
rate: 1 MHz. Rise time: 1 ns or less.
Flatness: ±2%.
Signal source for gain
and transient response.
TEKTRONIX PG 506
Calibration Generator
2. Leveled Sine-Wave
Generator
Frequency: 250 kHz to above
100 MHz. Output amplitude: variable
from 10 mV to 5 V p-p. Output
impedance: 50 Q. Reference fre-
quehcy: 50 kHz. Amplitude accuracy:
constant within 3% of reference fre
quency as output frequency changes.
Vertical, horizontal, and
triggering checks and
adjustments. Display
adjustments and Z-
Axis check.
TEKTRONIX SG 503
Leveled Sine-Wave
Generator.®
3. Time-Mark
Generator
Marker outputs: 10 ns to 0.5 s.
Marker accuracy: ±0.1%. Trigger
output: 1 ms to 0.1 ns, time-
coincident with markers.
Horizontal checks and
adjustments. Display
adjustment.
TEKTRONIX TG 501
Time-Mark Generator.®
4. Low-Frequency
Generator
Range: 1 kHz to 500 kHz. Output
amplitude: 300 mV. Output
impedance: 600 Q. Reference fre
quency: constant within 0.3 dB of
reference frequency as output fre
quency changes.
Low-frequency trigger
checks.
TEKTRONIX SG 502
Oscillator.®
5. Pulse Generator
Repetition rate: 1 kHz. Output ampli
tude: 5 V.
External clock and
storage checks.
TEKTRONIX PG 501
Pulse Generator.®
6. Test Oscilloscope
with 10X Probes
Bandwidth: dc to 100 MHz. Minimum
deflection factor: 5 mV/div. Accuracy:
+ 3%.
General troubleshoot
ing, holdoff check.
TEKTRONIX 2235
Oscilloscope.
7. Digital Voltmeter
(DMM)
Range: 0 to 140 V. Dc voltage accu
racy: ±0.15%. 4 1/2 digit display.
Power supply checks
and adjustments. Verti
cal adjustment.
TEKTRONIX DM 501A
Digital Multimeter
8. Coaxial Cable
(2 required)
Impedance: 50 Si. Length: 42 in. Con
nectors: BNC.
Signal interconnection.
Tektronix Part Number
012-0057-01.
9. Dual-Input Coupler Connectors: BNC female-to-dual-BNC
male.
Signal interconnection.
Tektronix Part Number
067-0525-01.
10. Coupler Connectors: BNC female-to-BNC
female.
Signal interconnection.
Tektronix Part Number
103-0028-00.
11. T-Connector Connectors: BNC. Signal interconnection.
Tektronix Part Number
103-0030-00.
12. Termination Impedance: 50 S2. Connectors: BNC.
Signal termination.
Tektronix Part Number
011-0049-01.
'Requires a TM 500-Series Power Module.
4-2

Table of Contents

Other manuals for Tektronix 2230

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2230 and is the answer not in the manual?

Tektronix 2230 Specifications

General IconGeneral
BrandTektronix
Model2230
CategoryTest Equipment
LanguageEnglish

Related product manuals