Performance Check Procedure— 2230 Service
j. Set the generator to produce a 0.1 s period, 2 ms
pulse duration, 5-division display.
n. Repeat parts c and d.
k. Repeat parts c through h.
o. Disconnect the test equipment from the instrument.
1. Set:
A SEC/DIV 1 ms
16. Check Average Mode
a. Set the WAVEFORM REFERENCE/MENU SELECT
switch to MENU SELECT (button out).
STORE/NON STORE NON STORE (button out)
ROLUSCAN SCAN (button out)
b. Use the Menu controls to select SWP LIMIT.
m. Set the generator to produce a 1 ms period, 20 *tS
pulse duration, 5-division display.
c. CHECK—The SWP LIMIT is adjustable from 1 to
2047 or NO LIMIT by rotating the CURSORS control.
4-11