Operating Basics
48 AFG3000 Series Quick Start User Manual
Set up Load Impedance
To set the load impedance, use the Output menu.
Quick Tips
The load impedance is applied to the amplitude, offset and high/low level settings.
When dBm is currently specified for output amplitude units, the amplitude units setting is automatically changed to V
pp
if you select high impedance.
Output impedance is always set to 50 .
1. Push the front-panel Sine > Continu-
ous button to display the Sine screen
in this example.
2. Push the front-panel Top Menu
button, and then push the Output
Menu bezel button.
3. The Output Menu is displayed.
4. To set the load impedance, push Load
Impedance.
5. The Load Impedance submenu is
displayed.
6. To adjust the load impedance, select
Load. You can set the load impedance
to any value from 1 to 10 k. The
default is 50 .
1
2
3
4
5
6