Trigger Setup
Push a side-be
zel button to select
sequencing the B trigger after the A a s Time
or as Events.
Time
(a) 8 ns
B Events
1
Set to
Minimum
5. Set the other Sequence Trigger parameters
in the related side- and lower-bezel menus.
B Trigger After Delay Time
The A trigger arms the instrument.
Posttrigg
er acquisition starts on the rst B
edge after the trigger delay time.
Trigge
r on B Events
The A tr
igger arms the instrument.
Posttrigger acquisition starts on the nth B
event.
Quick Tips
B-trigger delay time and horizontal position are independent functions. When you establish a trigger condition using
eit
her the A trigger alone or the A and B triggers together, you can also use the horizontal position control to delay the
acquisition by an additional amount.
When using the B trigger, the A and B trigger types can only be Edge.