This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument
complies.
EMC Compliance
EC Declarati
on of Conformity – EMC
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following
specifications as listed in the Official Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment for measurement, control, and
laborator
yuse.
1234
CISPR 11 :2
003. Radiated and conducted emissions, Group 1, Class A
IEC 61000
-4-2:2001. Electrostatic discharge immunity
IEC 61000
-4-3:2002. RF electromagnetic fi eld immunity
5
IEC 61000
-4-4:2004. Electrical fast transient/burst immunity
IEC 6100
0-4-5:2001. Power line surge immunity
IEC 6100
0-4-6:2003. Conducted RF immunity
6
IEC 6100
0-4-11:2004. Voltage dips and interruptions immunity
7
EN 61000-3-2:2006. AC power line harmonic emissions
EN 6100
0-3-3:1995.
Voltage changes, fluctuations, and flicker
European contact.
Tektronix UK, Ltd.
Western Peninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
This
product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur when this equipment is connected to a test object.
3
To ensure compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
Instrument rebooting may be experienced where the EUT takes longer than 10 seconds to recover from a transient i mmunity test.
5
The instrument will exhibit ≤ 4.0 division waveform displacement and ≤ 8.0 division increase in peak-to-peak noise when subjected
to radiated interference per IEC 61000-4-3.
6
The instrument will exhibit ≤ 1.0 division waveform displacement and ≤ 2.0 division increase in peak-to-peak noise when subjected
to conducted interference per IEC 61000-4-6.
7
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels