Analyze Wavefor
morTraceData
R1
3. Push Vertical in the side-bezel menu and
use the multipurpose knobs to adjust the
vertical sett
ings of the reference waveform
or trace.
Vertical
0.00 div
100 mV/div
4. Push Horizontal in the side-bezel menu
andusethemu
ltipurpose knobs to adjust
the horizontal settings of the reference
waveform or trace.
Horizontal
0.00 s
4.00 μs/div
5. Push Edit La
bel and use the resulting
menus to define labels to display with the
reference waveforms and traces.
Edit Labels
6. Pus Ref Details to read information
about the selected reference. Use this
to determ
ine whether the reference is an
analog waveform or an RF trace.
Ref Details
7. Push Sav
e to File to store reference
information in e xternal storage.
Save to
File
Quick Tips
Selecting and Displaying Reference Waveforms. You can display all the reference waveforms at the same time.
Push the appropriate screen button to select a particular reference waveform.
Removing Reference Waveforms from the Display. To remove a reference w aveform from the display, push the
front-panel R button to access the lower-bezel menu. Then push the associated button from the lower-bezel menu
to turn it off.
Scaling and Positioning a Reference Waveform. You can position and scale a reference w aveform independently
from all other displayed waveforms. Select the reference waveform and then adjust it with a multipurpose knob. You can
do this whether acquisition is running or not.
If a reference w aveform is selected, scaling and repositioning of the reference waveform operates the same way
whether zoom is turned on or off.
Saving 10 M and 20 M Reference Waveforms. 10 M and 20 M reference waveforms are volatile and not saved when
the oscilloscope power is turned off. To keep these waveforms, save them to external storage.