Appendix A: MDO4
000 Specifications
Table 1: Analog channel input and vertical specifications (cont.)
Characteristic Description
The limits stated below are for ambient temperature of ≤30 °C (86 °F) and the bandwidth selection
set to F ULL. Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Instrument
Volts/Div setting
Bandwidth
5 mV/div – 10 V/div DC to 500 MHz
2 mV/div – 4.98 mV/div DC to 350 MHz
MDO4104-3,
MDO4104-6
1 mV/div – 1.99 mV/div DC to 175 MHz
5 mV/div – 10 V/div DC to 500 MHz
2 mV/div – 4.98 mV/div DC to 350 MHz
Analog
Bandwidth, 1
MΩ,
DC coupled
MDO4054-3,
MDO4054-6
1 mV/div – 1.99 mV/div DC to 175 MHz
RMS noise (mV)
Instrument Bandwidth selection
1MΩ 50 Ω
Full Bandwidth
≤ (300 μV + 8.0% of
Volts/div setting)
≤ (75 μV + 6.0% of
Volts/div setting)
250 MHz bandwidth
limit
≤ (100 μV + 5.0% of
Volts/div setting)
≤ (50 μV + 4.0% of
Volts/div setting)
MDO4104-3,
MDO4104-6
20 MHz bandwidth
limit
≤ (100 μV + 5.0% of
Volts/div setting)
≤ (50 μV + 4.0% of
Volts/div setting)
Full Bandwidth
≤ (130 μV + 8.0% of
Volts/div setting)
≤ (130 μV + 8.0% of
Volts/div setting)
250 MHz bandwidth
limit
≤ (100 μV + 6.0% of
Volts/div setting)
≤ (100 μV + 6.0% of
Volts/div setting)
Random
Noise,
Sample Acquisition
Mode
MDO4054-3,
MDO4054-6
20 MHz bandwidth
limit
≤ (100 μV + 4.0% of
Volts/div setting)
≤ (100 μV + 4.0% of
Volts/div setting)
Table 2: Horizontal and acquisition system specifications
Characteristic Description
Long-term sample rate and
delay time accuracy
±5 ppm over any ≥1mstimeinterval
MDO4000 Series Oscilloscopes User Manual 181