Installation
Acquisition System: 1 M
The maximum input voltage at the BNC, 300 V
RMS
. Installation Category II.
Derate at 20 dB
/decade between 4.5 MHz and 45 MHz.
Derate at 14 dB/decade between 45 MHz and 450 MHz.
Above 450 MHz, 5 V
RMS
.
Acquisition System: 50
The maximum
input voltage at the BNC: 5 V
RMS
, with peaks ±20 V (DF 6.25%)
P6616: Digital Probe Inputs
The absolut
e maximum input voltage: ±42 V
Peak
.
Dedicated RF Input:
The maximu
m operating voltage: ±40 V
DC
.